ProductsDisposables > µ-Slide III 0.1

products
applications
references
press

contact





µ-Slide III 0.1

µ-Slide III0.1

3 assays on one µ-Slide for perfusion with rare samples
such as cells or blood

Instructions (PDF, 247 KB)
Shear stress and shear rates (PDF, 699 KB)


Applications

  • mouse blood experiments
  • rolling and adhesion assay
  • long term cultures of cells under perfusion
 

Technical features

  • channel cross section of 0.1 mm (height) x 1 mm (width)
  • small sample amounts
  • 1 mm scale for sample location
  • low dead volume
  • 3 parallel assays on one slide
  • also available as flow kit



dyne/cm² µl/min
1 9.37
10 93.69
100 936.94

Low flowrate for
physiological shear stress
Example experiment and reagent consumption

468 µl sample consumption only, in a one-way setup
for following experimental parameters:

  • shear stress 10 dyne / cm²
  • duration 5 min = 300 sec
  • viscosity 0.01 (dyne sec) / cm²

2 ml minimal sample volume for circular flow setup
using the ibidi perfusion system 

 

  technical details:
adapters female Luer
recommended volume per resservoir 60µl
number of channels 3
volume of each channel 4.5µl
height of channels 0.1mm
length of channels 45mm
width of channels 1mm
growth area per channel 0.43cm²
distance of scale bars 1mm
catalog numbers:
coated (Collagen IV) / 15 pcs. 80332
coated (Fibronectin)* / 15 pcs. 80333
hydrophobic, uncoated / 15 pcs. 80331

* available on request only

µ-Slide III 0.1